EMGA-930
Home > Products > CHNOS Analyzer
EMGA-930 O/N/H analyzer: It's a "New flagship model” for O/N/H analyzers and user friendly product.
This is a simultaneous oxygen, nitrogen & hydrogen elemental analyzer with high accuracy and repeatability suiting to cutting-edge technology’s R&D as well as quality control in the market of steel, new materials, catalyst and so on. This is a new generation model optimized to fit to user’s requests. The oxygen is measured as carbon monoxide and carbon dioxide by two non-dispersive infrared detectors, the nitrogen by a thermal conductivity detector and the hydrogen with non dispersive Infra Red detector as H2O.
Features
- Principle and Gas flow diagram
The sample is placed in a graphite crucible and the crucible is maintained between the upper and the lower electrode of the impulse furnace. A high current power passes through the crucible to create a high temperature.
The oxide in the sample reacts with the graphite crucible, and is extracted as carbon monoxide (CO), and carried together with the carrier gas. The extracted gas is directly analysed after the dust filter. According to the oxygen concentration, it is determined by CO or CO2 after pass throw oxidizer (Copper Oxide) with a non dispersive Infra-Red analyser (NDIR). The Hydrogen is determined with a non dispersive Infra Red analyser as H2O after H2 passes through the oxidizer (Copper Oxide).
- Super High Performance
*Range: Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25%
*Accuracy: sn-1?0.3ppm or CV?1.0% for STD sample sn-1?0.02ppm or CV?0.5% for Reference gas
*Accuracy H: sn-1?0.04ppm or RSD?2.0% for Reference gas
Double detectors of oxygen (CO and CO2), long type cell and optimized design of TCD provide wide range measurements.
Optimization of the signal processing reduces by half the electrical noise compared to conventional model.
- Measurement Window
Simple software allows easy handling. Extracted gas signals are displayed in real time numerically and also as curves together with temperature level. Graphs are saved automatically.
In the measurement window, sample weight can be registered automatically. Results are saved in a data table for easy management.
- HORIBA originality-Maintenance navigator
Maintenance counter informs users at the timing of consumables replacement to keep high accuracy means. And as navigator describes the easy to understand procedure for the replacing parts, operator can do maintenance works with absolutely no experience and technical knowledge.
Specifications
- EMGA-930: Oxygen/Nitrogen/Hydrogen analyzer
- Principle: Oxygen: Non Dispersive Infrared detector Nitrogen: Thermal Conductivity detector(TCD) Hydrogen: Non dispersive Infrared detector (as H2O)
- Measurement range*: Oxygen: 0-5%, Nitrogen: 0-3%, Hydrogen: 0-0.25% *Up to 100% wt is possible by decreasing the sample weight.
- Sample weight: 1.0 ± 0.1g
- Sensitivity(Minimum reading): Oxygen/Nitrogen/Hydrogen: 0.001 ppm.
- Accuracy**(Repeatability): Oxygen/Nitrogen σn-1≦0.02 ppm or RSD≦0.5% whichever is larger (Reference gas) σn-1≦0.3 ppm or RSD≦1.0% whichever is larger (Standard sample) Hydrogen σn-1≦0.04 ppm or RSD≦2% whichever is larger (Reference gas) ** Based on the sample weight 1.0 ± 0.1g and with automation functions described below
- Display: Measurement result: LCD or printout Alarm message: LCD or printout Flow sheet: LCD
- Power of furnace: Inert gas fusion in impulse furnace Enable to set the generating power 0 – 8.0 kw
- Sample introduction: Dual sample/flux introduction mechanism
- Automation functions: Auto cleaner, Crucible loader